FUNDAMENTALS OF UV MEASUREMENT
 

Peak Intensity Variances With Speed

The variance when measuring peak intensity (mW/cm2) lies in the sample rate of the radiometer versus the time spent under the focal point of the lamp (speed).

A focused UV lamp, has a 3/4 inch focal plane. The UV Power Puck® samples at 25 samples per second. Using the sample rate of the Power Puck, we find that at the 10m/min speed, it is able to accumulate 3.85 samples each inch, but only 2.88 samples which are under the focal plane.

For example:

10m/min = 390 inches/minute
= 6.5 inches/second
25 samples/sec = 25 samples/6.5 inches
= 3.85 samples/inch
.75 inch in the focal plane = 2.88 samples under the focal plane (3.85 X .75)

Using the above math, here is what happens at 20m/min:

20m/min = 780 inches/minutes
= 13 inches/second

Now:

25 samples/second (capability of UV Power Puck) = 25 samples/13 inches

Or:

but = 1.92 samples per inch
= only 1.44 samples under the focus (1.92 X .75)

At some other speeds, here are the results:

  • 30m/min = .96 samples under the focus
  • 40m/min = .72 samples under the focus
  • 50m/min = .58 samples under the focus
  • 60m/min = .48 samples under the focus
  • 70m/min = .41 samples under the focus
So you can see any speed over 20m/min has less than 1 chance in 25 of sampling under the focus of the lamp.

**Note -- To match a UVIMAP more closely with UV Power Puck or UVICURE Plus, set the UVIMAP sampling rate at 20 samples/sec.


EIT, Inc., 108 Carpenter Drive, Sterling, VA 20164
Phone: 703-478-0700 · Fax: 703-478-0815

Questions, problems, comments? Contact webmaster@eitinc.com
All contents © 1998 EIT Inc. except where noted. All rights reserved.

Last updated