UVIRAD Operating Instructions


DO NOT USE WHERE ENERGY LEVELS EXCEED 100mW/cm2

Important: UVIRAD should be charged overnight before using for the first time.

I. OPERATING INSTRUCTIONS

  1. Remove UVIRAD from battery charger.

  2. Press "ON/RESET" button. Display will come on displaying all zeros.

  3. Place UVIRAD in UV environment to be measured.

  4. Read total accumulated energy level on display.

  5. To clear display, press "ON/RESET" button.

  6. UVIRAD is now ready to make another reading.

  7. Battery charging instructions.

II. INDICATORS

Low Battery: Display will flash ON and OFF when the level of charge is too low for reliable operation.

III. "ON/RESET"

The unit can be turned on or reset by pressing the "ON/RESET" button. It will remain on for 45 seconds after its last exposure to ultraviolet levels of at least 100mW/cm2. Unit will then turn itself OFF to conserve batteries.

IV. CLEANING

Clean unit only with denatured alcohol and a clean cloth or cotton swab. To ensure accuracy, clean detector window periodically with denatured alcohol and a cotton swab. Avoid touching detector window with hands or fingers.

V. WARRANTY

Warranty period for UVIRAD is 6 months from the date of purchase, abuse and neglect excepted. There are no user serviceable parts inside UVIRAD. If your unit does not operate, contact EIT for service. We will arrange for its return to the factory for repair and recalibration.

* Specifications are subject to change. EIT will not be responsible for discrepancies due to typographic errors.


 
Return to Technical Product Notes


EIT, Inc., 108 Carpenter Drive, Sterling, VA 20164
Phone: 703-478-0700 · Fax: 703-478-0815

Questions, problems, comments? Contact webmaster@eitinc.com
All contents © 1998 EIT Inc. except where noted. All rights reserved.

Last updated